Technical Specifications
Cond. Measuring Range
Conductivity:
0~200 mS/cm, divided into five ranges, automatically switch measuring range.
(0 to 19.99) μS/cm;
(20.0 to 199.9) μS/cm;
(200 to 1999) μS/cm;
(2.00 to 19.99) mS/cm;
(20.0 to 199.9) mS/cm
TDS: (0 to 100) ppt(g/L)
Resolution :
0.01/0.1/1μS/cm 0.01/0.1 mS/cm
Accuracy :
±1.0% F.S.
Temperature Compensation Range :
(0 to 50)˚C (Auto. or Manual)
Cell constant :
0.1/1/10 cm-1
Temp.
Measuring Range : 0 to 100℃
Resolution : 0.1℃
Accuracy : ±0.5℃±1 digit
Other Specifications
Data Storage : N/A
Storage Content : N/A
Power : 3* AA alkaline batteries
Dimension & Weight :
Meter: (88*170*33)mm / 313g
Case: (360*270*76)mm / 1.6kg
Connector : RCA for temperature, 8-pin for conductivity
What’s in the box? : EC850 meter, 2301T-S conductivity/temp. electrode, 50ml of conductivity standards (1413µS, 12.88mS, 84µS), screwdriver (for battery replacement), instruction manual, carrying case